Summary

Successful exploitation of this vulnerability in a custom integration version could allow an attacker to steal an authenticated clinician’s token via a crafted link.

The following versions of OHIF Viewers DICOM are affected:

  • OHIF DICOM Web Viewer Framework <=v3.12.0
CVSS Vendor Equipment Vulnerabilities
v3 8.2 Open Health Imaging Foundation (OHIF) OHIF Viewers DICOM Server-Side Request Forgery (SSRF)

Background

  • Critical Infrastructure Sectors: Healthcare and Public Health
  • Countries/Areas Deployed: Worldwide
  • Company Headquarters Location: United States

Vulnerabilities

Expand All +

CVE-2026-12473

Two data sources (DICOMWebProxy and DICOMJSON) shipped in the default configuration fetch an arbitrary URL parameter without validation. A global authentication service in OHIF automatically injects the authenticated user’s OIDC Bearer token into the resulting requests, sending it to the attacker-controlled server. DICOMweb data sources are not impacted.

View CVE Details


Affected Products

OHIF Viewers DICOM
Vendor:
Open Health Imaging Foundation (OHIF)
Product Version:
Open Health Imaging Foundation (OHIF) OHIF DICOM Web Viewer Framework: <=v3.12.0
Product Status:
known_affected
Remediations

Mitigation
The maintainer has fixed the reported vulnerability and released version 3.12.2 (2026-05-18). The fix is located at OHIF/Viewers#5985 (master), OHIF/Viewers#5978 (release/3.12).

Mitigation
Users are recommended to upgrade to v3.12.2 or later. Operators who need dicomwebproxy or dicomjson in authenticated deployments must additionally configure the new dangerouslyAllowedOriginsForAuthenticatedEnvironments allowlist in app-config.js.

Mitigation
Users running OHIF with authentication should remove ALL unused DicomWebProxyDataSource and DicomJSONDataSource configurations from the configuration file they are deploying with.

Relevant CWE: CWE-918 Server-Side Request Forgery (SSRF)


Metrics

CVSS Version Base Score Base Severity Vector String
3.1 8.2 HIGH CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N
4.0 8.3 HIGH CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N

Acknowledgments

  • Simon Weber and Volker Schönefeld of Machine Spirits UG reported this vulnerability to CISA

Legal Notice and Terms of Use

This product is provided subject to this Notification (https://www.cisa.gov/notification) and this Privacy & Use policy (https://www.cisa.gov/privacy-policy).


Recommended Practices

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability.

Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.

Locate control system networks and remote devices behind firewalls and isolating them from business networks.

When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations

[…]