ZOLL ePCR IOS Mobile Application

View CSAF

Summary

Successful exploitation of this vulnerability could allow an attacker to gain unauthorized access to protected health information (PHI) or device telemetry.

The following versions of ZOLL ePCR IOS Mobile Application are affected:

  • ePCR IOS Mobile Application 2.6.7 (CVE-2025-12699)
CVSS Vendor Equipment Vulnerabilities
v3 5.5 ZOLL ZOLL ePCR IOS Mobile Application Insertion of Sensitive Information into Externally-Accessible File or Directory

Background

  • Critical Infrastructure Sectors: Healthcare and Public Health
  • Countries/Areas Deployed: Worldwide
  • Company Headquarters Location: United States

Vulnerabilities

Expand All +

CVE-2025-12699

The ZOLL ePCR IOS application reflects unsanitized user input into a WebView. Attacker-controlled strings placed into PCR fields (run number, incident, call sign, notes) are interpreted as HTML/JS when the app prints or renders that content. In the proof of concept (POC), injected scripts return local file content, which would allow arbitrary local file reads from the app’s runtime context. These local files contain device and user data within the ePCR medical application, and if exposed, would allow an attacker to access protected health information (PHI) or device telemetry.

View CVE Details


Affected Products

ZOLL ePCR IOS Mobile Application
Vendor:
ZOLL
Product Version:
ZOLL ePCR IOS Mobile Application: 2.6.7
Product Status:
known_affected
Remediations

Mitigation
ZOLL ePCR IOS application was decommissioned in May 2025. ZOLL has no current plans to provide a replacement application. If users have questions or concerns, they are encouraged to reach out directly to ZOLL Support. https://www.zolldata.com/contact-us.

Relevant CWE: CWE-538 Insertion of Sensitive Information into Externally-Accessible File or Directory


Metrics

CVSS Version Base Score Base Severity Vector String
3.1 5.5 MEDIUM CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N

Acknowledgments

  • Bryan Riggins reported this vulnerability to ZOLL

Legal Notice and Terms of Use

This product is provided subject to this Notification (https://www.cisa.gov/notification) and this Privacy & Use policy (https://www.cisa.gov/privacy-policy).


Recommended Practices

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:

Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.

Locate control system networks and remote devices behind firewalls and isolating them from business networks.

When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implem

[…]
Content was cut in order to protect the source.Please visit the source for the rest of the article.

This article has been indexed from All CISA Advisories

Read the original article: